108AA Single Point Load Cell

IMAGES
NTEP OIML

The Anyload 108AA is a compact aluminum single-point load cell for ultra-low-capacity applications, available from 100 g to 5 kg with IP66 protection and with CE, ATEX, FM, IECEx, and RoHS certifications.
OIML C3 certified for 1 to 5 kg capacity range, its small lightweight design is ideal for precision OEM weighing instruments, laboratory scales, and light-duty bench scales.

PRODUCT DETAILS

The 108AA is a compact, reliable aluminum single-point load cell built for the lightest end of the weighing spectrum — from 100 grams to 5 kgs. Designed for OEM integration into precision instruments, it features IP66 environmental sealing and a full-scale output of 1.0 mV/V (≤ 500 g) or 2.0 mV/V ( ≥1 kg), optimized for high-accuracy signal conditioning electronics. Platform sizes are 200×200 mm or 7.87 x 7.87 inches for capacities up to 500 g, and 230×230 mm or 9 x 9 inches for 1 kg to 5 kg. A compensated temperature range of −10°C to +40°C ensures stable measurement across typical indoor operating conditions. CE, RoHS, ATEX, FM, and IECEx certifications support global deployment in both standard and hazardous environment.

Most Common Applications: Laboratory balances, precision OEM scales, jewelry scales, postal scales, micro-dosing and dispensing systems, bench scales.

  • Ultra-Low Capacity Range (100 g to 5 kg): Covers the lightest weighing needs for laboratory instruments, micro-dosing devices, and precision OEM applications, with OIML C3 certification for the 1 to 5 kg range.
  • Output Rating: 1.0 mV/V full-scale output for capacities ≤ 500 g and 2.0 mV/V for capacities ≥1 kg, allowing optimized signal conditioning across the full range.
  • IP66 Protection Rating: Sealed against dust and high-pressure water jets, providing reliable performance in light industrial and commercial environments.
  • ATEX, FM & IECEx Certified: .
  • ATEX certified for a single load cell to II 1 G Ex ia IIC T6 Ga; -40°C < Ta < +60°C & II 1 D Ex ia IIIC T83°C Da; -40°C < Ta < +60°C. For multiple load cells to II 1 G Ex ia IIC T4 Ga; -40°C < Ta < +60°C FM certified for a single load cell to Class I, II and III, Division 1, Groups A, B, C, D, E, F and G; T6; -40°C < Ta < +60°C. For multiple load cells to Class I, Division 1, Groups A, B, C and D; T4; -40°C < Ta < +60°C IECEx certified for a single load cell to Ex ia IIC T4 Ga.

  • 150% Safe Overload: Built-in overload protection guards the sensor against accidental overloading — essential in low-capacity precision applications where excess force can easily damage sensitive components.
Capacities

100 g (0.22 lb), 200 g (0.44 lb), 300 g (0.66 lb), 500 g (1.10 lb), 1 kg (2.20 lb), 2 kg (4.41 lb), 3 kg (6.61 lb), 5 kg (11.02 lb)

Combined Error

Non-Linearity

<±0.023% FS (≤500 g); <±0.017% FS (≥1 kg)

Safe Overload

150% of full scale

Ultimate Overload

300% of full scale

Full Scale Output

1.0 mV/V ±20% (≤500 g); 2.0 mV/V ±20% (≥1 kg)

Maximum Excitation

15 V DC

Bridge Resistance

350 Ω ±3

Operating Temperature

−18°C to +65°C (−0.4°F to +149°F)

Compensated Temperature

−10°C to +40°C (14°F to +104°F)

Sealing Standard

IP66

Material

Aluminum

Protective Coating